Títol: Verifcation of dose calculations in a heterogeneous phantom using PRIMO
Autor: Ortiz Catalán, Ramon
Director/Tutor: Fernández Varea, José María
Matèria: Dosimetria (Radiació)
Mètode de Montecarlo
Treballs de fi de grau
Radiation dosimetry
Monte Carlo method
Bachelor's theses
Data de publicació: gen-2018
Resum: Treatment planning systems (TPSs) are routinely used in radiotherapy to calculate the monitor units (MU) required to deposit a certain amount of energy per unit mass in the patient and the dose distribution in it due to their high calculation speed. To reduce calculation time, TPSs adopt several approximations which are optimized for low atomic number materials such as those found in the human body. These approximations may lead to substantial deviations between calculated and measured dose in the presence of heterogeneities. In the present TFG we employ Monte Carlo simulations to study these discrepancies and verify the TPS calculations. The program used is PRIMO, which is based on the penelope code system.
Nota: Treballs Finals de Grau de Física, Facultat de Física, Universitat de Barcelona, Curs: 2018, Tutor: José M. Fernández-Varea