dc.description.abstractBackground: In recent years, subcutaneous implantable cardioverter-defibrillator (S-ICD) implants have progressively increased and have been shown to be safe and highly successful, affording low reintervention rates regardless of the technique used. Case summary: We present a case of S-ICD implantation in a patient diagnosed with idiopathic ventricular fibrillation. In the first follow-up consultation the patient showed appropriate detection parameters in the three configurations. However, chest X-ray revealed lead displacement with a tip migration from the manubrium area of the sternum to the xiphoid process. Discussion: This case highlights the importance of performing at least one chest X-ray during the first weeks after S-ICD implantation, allowing the detection of a problem such as lead displacement, which can lead to undersensing of ventricular arrhythmias or S-ICD oversensing.-
dc.format.extent5 p.-
dc.format.mimetypeapplication/pdf-
dc.language.isoeng-
dc.publisherOxford University Press-
dc.relation.isformatofReproducció del document publicat a: https://doi.org/10.1093/ehjcr/ytz189-
dc.relation.ispartofEuropean Heart Journal - Case Reports, 2019, vol. 3, num. 4, p. 1-5-
dc.relation.urihttps://doi.org/10.1093/ehjcr/ytz189-
dc.rightscc-by-nc (c) Apolo, José et al., 2019-
dc.rights.urihttp://creativecommons.org/licenses/by-nc/3.0/es-
dc.sourceArticles publicats en revistes (Medicina)-
dc.subject.classificationRaigs X-
dc.subject.classificationDesfibril·ladors cardioversors implantables-
dc.subject.classificationFibril·lació ventricular-
dc.subject.otherX-rays-
dc.subject.otherImplantable cardioverter-defibrillators-
dc.subject.otherVentricular fibrillation-
dc.titleUndetected displacement of a subcutaneous implantable cardioverter-defibrillator lead-
dc.typeinfo:eu-repo/semantics/article-
dc.typeinfo:eu-repo/semantics/publishedVersion-
dc.identifier.idgrec698389-
dc.date.updated2020-06-01T21:16:23Z-
dc.rights.accessRightsinfo:eu-repo/semantics/openAccess-
Appears in Collections:Articles publicats en revistes (IDIBAPS: Institut d'investigacions Biomèdiques August Pi i Sunyer)
Articles publicats en revistes (Medicina)

Files in This Item:
File Description SizeFormat 
698389.pdf538.16 kBAdobe PDFView/Open


This item is licensed under a Creative Commons License Creative Commons