Title: Ultra-high-performance liquid chromatography-atmospheric pressure ionization-tandem mass spectrometry method for the migration studies of primary aromatic amines from food contact materials
Author: Arrizabalaga Larrañaga, Ane
de Juan-de Juan, Pedro
Bressan, Claudia
Vázquez-Espinosa, Mercedes
González-de-Peredo, Ana V.
Santos Vicente, Francisco Javier
Moyano Morcillo, Encarnación
Keywords: Espectrometria de masses
Cromatografia de líquids d'alta resolució
Ionització
Mass spectrometry
High performance liquid chromatography
Ionization
Issue Date: 31-Jan-2022
Publisher: Springer Verlag
Abstract: This work describes the development of an ultra-high-performance liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS) method for the determination of 23 primary aromatic amines (PAAs) that can potentially migrate from food contact materials. The chromatographic separation was performed in a pentafluorophenylpropyl (PFPP) column achieving the separation of all PAAs in less than 6.5 min using water to acetonitrile (0.1% acetic acid in both solvents) as mobile phase and a gradient elution. The feasibility of atmospheric pressure chemical ionization (APCI) and atmospheric pressure photoionization (APPI) was evaluated as alternative to electrospray ionization (ESI) for the analysis of PAAs. Results showed that for most of the compounds, better responses were obtained with APCI, which shows the advantage of being less susceptible to matrix effects. Tandem mass spectrometry (MS/MS) fragmentation studies of [M + H]+ allowed for the selection of the two most characteristic and abundant product ions of the 23 PAAs which led to the development of a selective and sensitive UHPLC-APCI-MS/MS method with limits of detection ranging from 0.2 to 2 μg kg− 1. Moreover, intraday and inter-day precisions of the method in terms of relative standard deviation (RSD%) were lower than 10% and 15%, while trueness as relative error was <15% for most of the compounds. The UHPLC-APCI-MS/MS method was applied to the analysis of twenty black Nylon kitchenware samples that were submitted to migration tests using food simulant B (3% acetic acid, w/v), and the presence of PAAs were detected in eighteen samples at concentrations above the legislated limit (2 μg kg−1 of food or food simulants).
Note: Reproducció del document publicat a: https://doi.org/10.1007/s00216-022-03946-3
It is part of: Analytical and Bioanalytical Chemistry, 2022, vol. 414, p. 3137-3151
URI: http://hdl.handle.net/2445/194669
Related resource: https://doi.org/10.1007/s00216-022-03946-3
ISSN: 1618-2642
Appears in Collections:Articles publicats en revistes (Enginyeria Química i Química Analítica)

Files in This Item:
File Description SizeFormat 
722255.pdf1 MBAdobe PDFView/Open