Please use this identifier to cite or link to this item: https://hdl.handle.net/2445/182277
Title: Vibrating boron-doped diamond electrode: a new, durable and highly sensitive tool for the detection of cadmium
Author: Morris, Adam
Serrano i Plana, Núria
Díaz-Cruz, José Manuel
Bendavid, Avi
Chen, Miao
Vepsäläinen, Mikko
Keywords: Elèctrodes
Cadmi
Voltametria
Electrodes
Cadmium
Voltammetry
Issue Date: 15-Dec-2021
Publisher: Elsevier B.V.
Abstract: In this paper, a vibrating boron-doped diamond (BDD) electrode electroanalytical device and respective method for the analysis of ultralow concentrations of Cd(II) in water were studied. The enhanced mass transfer on the electrode surface was studied using Ru(NH3)6Cl3. Vibration with 133 Hz frequency enhanced the Ru(III) to Ru(II) reduction by 92.6% compared to a static electrode. The peak current of the anodic stripping voltammetry (ASV) method employed was increased by a factor of 5.3 and 4.7 for 10 and 30 microg L-1 Cd(II) concentrations, respectively, when a frequency of 200 Hz was used. A calibration plot with two linear regions was resolved between 0.01 and 1 microg L-1 and 1-30 microg L-1 with the LOD and LOQ of 0.04 microg L-1 and 0.12 microg L-1, respectively. The applicability of the device and the respective method in the analysis of real environmental samples was successfully verified by analysis of river samples and comparing the results with the ICP analysis presenting high reproducibility and trueness. According to the results of this research, the vibrating BDD electrode with the ASV method has excellent analytical performance without surface modification or regular replacement or polishing of the electrode surface. Combining the exceptional electrochemical and chemical properties of BDD with enhanced mass transfer and signal strength of vibrating electrodes makes the system especially suitable for on-site and online analysis of heavy metals.
Note: Reproducció del document publicat a: https://doi.org/10.1016/j.aca.2021.339166
It is part of: Analytica Chimica Acta, 2021, vol. 1188, p. 339166
URI: https://hdl.handle.net/2445/182277
Related resource: https://doi.org/10.1016/j.aca.2021.339166
ISSN: 0003-2670
Appears in Collections:Articles publicats en revistes (Enginyeria Química i Química Analítica)

Files in This Item:
File Description SizeFormat 
715334.pdf1.02 MBAdobe PDFView/Open


This item is licensed under a Creative Commons License Creative Commons