Please use this identifier to cite or link to this item: https://hdl.handle.net/2445/221512
Full metadata record
DC FieldValueLanguage
dc.contributor.authorRodríguez García, María-
dc.contributor.authorMartínez Martín, Irene-
dc.contributor.authorAliart, Irene-
dc.contributor.authorSainz de Medrano, Jaime I.-
dc.contributor.authorRico, Nayra-
dc.contributor.authorEscudero Saiz, Víctor Joaquín-
dc.contributor.authorMaduell, Francisco-
dc.contributor.authorMorales Ruiz, Manuel-
dc.contributor.authorCasals Mercadal, Gregori-
dc.date.accessioned2025-06-12T12:36:16Z-
dc.date.available2025-06-12T12:36:16Z-
dc.date.issued2025-02-08-
dc.identifier.issn1420-3049-
dc.identifier.urihttps://hdl.handle.net/2445/221512-
dc.description.abstractAccurate quantification of indoxyl sulfate (IndS) and p-cresyl sulfate (pCS) is essential for understanding their role in chronic kidney disease (CKD) progression and for developing strategies to mitigate their harmful effects, including cardiovascular morbidity and renal fibrosis. Advances in liquid chromatography-high-resolution mass spectrometry (LC-HRMS) enable the integration of powerful diagnostic tools into clinical laboratories. Along with accurate quantification, precise mass measurements allow for untargeted compound identification. Methods: An LC-HRMS was validated for quantifying IndS and pCS in human serum, following EMA guidelines. The method involved protein precipitation with methanol, micro-LC for chromatographic separation, and detection based on accurate mass, with simultaneous high-resolution full-scan acquisition. Clinical samples from patients with varying degrees of renal insufficiency and samples obtained before and after hemodiafiltration were analyzed. Results: The method demonstrated acceptable linearity, precision, and accuracy. The measurement range for both analytes was from 100 to 40,000 ng/mL. Serum levels of IndS and pCS correlated with decreased renal function. After hemodiafiltration, there was a significant reduction of IndS (50%) and pCS (43%). Simultaneous untargeted analysis allowed to identify metabolites significantly underexpressed after hemodiafiltration. Conclusions: An accurate LC-HRMS method was validated for the quantification of IndS and pCS serum levels in patients with CKD, providing insights into toxin dynamics and enabling untargeted metabolic evaluation.-
dc.format.extent14 p.-
dc.format.mimetypeapplication/pdf-
dc.language.isoeng-
dc.publisherMDPI-
dc.relation.isformatofReproducció del document publicat a: https://doi.org/10.3390/molecules30040782-
dc.relation.ispartofMolecules, 2025, vol. 30, num.4-
dc.relation.urihttps://doi.org/10.3390/molecules30040782-
dc.rightscc-by (c) Rodríguez-García, M. et al., 2025-
dc.rights.urihttp://creativecommons.org/licenses/by/4.0/-
dc.sourceArticles publicats en revistes (Biomedicina)-
dc.subject.classificationMetabolòmica-
dc.subject.classificationMalalties del ronyó-
dc.subject.classificationCromatografia de líquids d'alta resolució-
dc.subject.classificationEspectrometria de masses-
dc.subject.otherMetabolomics-
dc.subject.otherKidney diseases-
dc.subject.otherHigh performance liquid chromatography-
dc.subject.otherMass spectrometry-
dc.titleValidation of an LC–HRMS Method for Quantifying Indoxyl Sulfate and p-Cresyl Sulfate in Human Serum-
dc.typeinfo:eu-repo/semantics/article-
dc.typeinfo:eu-repo/semantics/publishedVersion-
dc.identifier.idgrec758785-
dc.date.updated2025-06-12T12:36:17Z-
dc.rights.accessRightsinfo:eu-repo/semantics/openAccess-
dc.identifier.pmid40005095-
Appears in Collections:Articles publicats en revistes (Biomedicina)
Articles publicats en revistes (IDIBAPS: Institut d'investigacions Biomèdiques August Pi i Sunyer)

Files in This Item:
File Description SizeFormat 
894562.pdf1.3 MBAdobe PDFView/Open


This item is licensed under a Creative Commons License Creative Commons