Please use this identifier to cite or link to this item: https://hdl.handle.net/2445/222115
Title: Toward Sub-100 ps TOF-PET Systems Employing the FastIC ASIC With Analog SiPMs
Author: Mariscal-Castilla, Antonio
Gomez Fernandez, Sergio
Manera Escalero, Rafel
Fernandez-Tenllado, José María
Mauricio Ferre, Joan
Kratochwil, Nicolaus
Alozy, J.
Piller, Markus
Portero, Sara
Sanuy Charles, Andreu
Guberman, Daniel
Silva, J. J.
Auffray, Etiennette
Ballabriga, Rafael
Ariño Estrada, Gerard
Campbell, Michael
Gascón Fora, David
Keywords: Tomografia per emissió de positrons
Sistemes d'imatges
Positron emission tomography
Imaging systems
Issue Date: 1-Sep-2024
Publisher: Institute of Electrical and Electronics Engineers (IEEE)
Abstract: Time of Flight positron emission tomography (TOFPET) scanners demand electronics that are power-efficient, low-noise, cost-effective, and possess a large bandwidth. Recent developments have demonstrated sub-100 ps time resolution with elevated power consumption per channel, rendering this unfeasible to build a scanner. In this work, we evaluate the performance for the TOF-PET of the FastIC front-end using different scintillators and silicon photomultipliers (SiPMs). FastIC is an eight-channel application specific integrated circuit developed in CMOS 65 nm capable of measuring the energy and the arrival time of a detected pulse with 12 mW per channel. Using Hamamatsu SiPMs (S13360-3050PE) coupled to LSO:Ce:0.2%Ca crystals of 2 × 2 × 3 mm3 and LYSO:Ce:0.2%Ca of 3.13 × 3.13 × 20 mm3, we measured a coincidence time resolution (CTR) of (95 ± 3) and 156 ± 4) ps full width half maximum (FWHM), respectively. With Fondazione Bruno Kessler NUV-HD LF2 M0 SiPMs coupled to the same crystals, we obtained a CTR of (76 ± 2) and (127 ± 3) ps FWHM. We employed FastIC with a TlCl pure Cherenkov emitter, demonstrating time resolutions comparable to those achieved with the high-power-consuming electronics. These findings shows that the FastIC represents a cost-effective alternative that can significantly enhance the time resolution of the current TOF-PET systems while maintaining low power consumption. Index Terms—Fast front-end electronics, photo-detector technology, positron emission tomography, scintillator, silicon photomultipliers (SiPMs), Time of Flight positron emission tomography (TOF-PET).
Note: Reproducció del document publicat a: https://doi.org/10.1109/TRPMS.2024.3414578
It is part of: IEEE Transactions on Radiation and Plasma Medical Sciences, 2024, vol. 8, num.7, p. 718-733
URI: https://hdl.handle.net/2445/222115
Related resource: https://doi.org/10.1109/TRPMS.2024.3414578
ISSN: 2469-7311
Appears in Collections:Articles publicats en revistes (Física Quàntica i Astrofísica)

Files in This Item:
File Description SizeFormat 
868347.pdf3.51 MBAdobe PDFView/Open


This item is licensed under a Creative Commons License Creative Commons