Please use this identifier to cite or link to this item: https://hdl.handle.net/2445/224255
Title: Agreement and reproducibility between 3DStent vs. Optical Coherence Tomography for evaluation of stent area and diameter
Author: Ruberti, Andrea
Rinaldi, Riccardo
Occhipinti, Giovanni
Ramus, Liliane
Guagliumi, Giulio
Sabaté Tenas, Manuel
Brugaletta, Salvatore
Keywords: Diagnòstic per la imatge
Cardiologia
Visualització tridimensional
Diagnostic imaging
Cardiology
Three-dimensional display systems
Issue Date: 28-Oct-2024
Publisher: Springer Verlag
Abstract: 3DStent is a novel rotational angiography imaging capable of 3D reconstruction and measuring stent area and diameter, without need for intravascular imaging. To compare 3DStent and OCT-derived stent area and diameter after PCI. Patients with de novo coronary lesions who underwent treatment with a single DES and evaluated by OCT and 3DStent were included. Stent area and diameter were measured by 3DStent, at abluminal, mid and endoluminal side and by OCT. From September 2023 to February 2024 six coronary lesions were analyzed. Post-PCI stent area measured by OCT was (mean +/- standard deviation) 7.03 +/- 2.85 mm(2) and by 3DStent 9.41 +/- 2.79 mm(2), 7.21 +/- 2.23 mm(2) and 5.63 +/- 1.83 mm(2) at abluminal, mid and endoluminal side, respectively. Stent diameter by OCT was 2.93 +/- 0.58 mm, and by 3DStent 3.27 +/- 0.50 mm, 2.86 +/- 0.49 mm and 2.52 +/- 0.45 mm at abluminal, mid and endoluminal side, respectively. Significant correlation was observed between OCT and 3DStent in relation to stent area (Exp(B) 3.35, mean of difference 0.19 +/- 1.01 mm(2), 95%CI -1.80-2.17 mm(2), p < 0.001) and diameter (Exp(B) 3.18, mean difference - 0.07 +/- 0.18 mm, 95%CI -0.43-0.30 mm, p < 0.001), particularly when 3DStent measurements performed at the mid side. Very high reproducibility was demonstrated by intra- and inter-observer analysis (r = 0.92 and r = 0.93 respectively). 3DStent appears to be an easy and reproducible tool to assess post-PCI stent area and diameter as compared to OCT.
Note: Reproducció del document publicat a: https://doi.org/10.1007/s10554-024-03268-8
It is part of: International Journal of Cardiovascular Imaging, 2024, vol. 40, p. 2581-2589
URI: https://hdl.handle.net/2445/224255
Related resource: https://doi.org/10.1007/s10554-024-03268-8
ISSN: 1569-5794
Appears in Collections:Articles publicats en revistes (Medicina)

Files in This Item:
File Description SizeFormat 
872793.pdf1.7 MBAdobe PDFView/Open


This item is licensed under a Creative Commons License Creative Commons