Please use this identifier to cite or link to this item: http://hdl.handle.net/2445/25623
Title: Performance of mechanical ventilators at the patient's home: a multicentric quality control study
Author: Farré Ventura, Ramon
Navajas Navarro, Daniel
Prats, E.
Marti, Sara
Guell, R.
Montserrat Canal, José Ma.
Tebé, Cristian
Escarrabill Sanglas, Joan
Keywords: Respiració artificial
Respiradors
Infermeria a domicili
Artificial respiration
Respirators
Home nursing
Issue Date: 2006
Publisher: BMJ Publishing Group Ltd
Abstract: Background: Quality control procedures vary considerably among the providers of equipment for home mechanical ventilation (HMV). Methods: A multicentre quality control survey of HMV was performed at the home of 300 patients included in the HMV programmes of four hospitals in Barcelona. It consisted of three steps: (1) the prescribed ventilation settings, the actual settings in the ventilator control panel, and the actual performance of the ventilator measured at home were compared; (2) the different ventilator alarms were tested; and (3) the effect of differences between the prescribed settings and the actual performance of the ventilator on non-programmed readmissions of the patient was determined. Results: Considerable differences were found between actual, set, and prescribed values of ventilator variables; these differences were similar in volume and pressure preset ventilators. The percentage of patients with a discrepancy between the prescribed and actual measured main ventilator variable (minute ventilation or inspiratory pressure) of more than 20% and 30% was 13% and 4%, respectively. The number of ventilators with built in alarms for power off, disconnection, or obstruction was 225, 280 and 157, respectively. These alarms did not work in two (0.9%), 52 (18.6%) and eight (5.1%) ventilators, respectively. The number of non-programmed hospital readmissions in the year before the study did not correlate with the index of ventilator error. Conclusions: This study illustrates the current limitations of the quality control of HMV and suggests that improvements should be made to ensure adequate ventilator settings and correct ventilator performance and ventilator alarm operation.
Note: Reproducció del document publicat a: http://dx.doi.org/10.1136/thx.2005.052647
It is part of: Thorax, 2006, vol. 61, num 5, p. 400-404
URI: http://hdl.handle.net/2445/25623
Related resource: http://dx.doi.org/10.1136/thx.2005.052647
ISSN: 0040-6376
Appears in Collections:Articles publicats en revistes (Ciències Fisiològiques)

Files in This Item:
File Description SizeFormat 
539157.pdf152.64 kBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.