dc.description.abstractThe usual setup for measuring respiratory input impedance requires a differential pressure transducer attached to a pneumotachograph. Because no data correction procedure has been devised to account for transducer asymmetry, a highly symmetrical transducer is required to obtain reliable impedance data. Here, a general model for the measuring system is presented. Its main feature is that differential pressure transducers are modeled as two-input-one-output systems. From the theoretical model, a dynamic calibration and data correction procedure is defined. This was tested using highly asymmetrical transducers (common-mode rejection ratio between 45 and 27 dB) to measure the impedance of two respiratory analogs. Results obtained show that respiratory input impedance can be adequately measured if data are corrected for transducer asymmetry.-
dc.format.extent4 p.cat
dc.format.mimetypeapplication/pdfeng
dc.language.isoengeng
dc.publisherIEEEcat
dc.relation.isformatofReproducció del document publicat a http://dx.doi.org/10.1109/10.40822cat
dc.relation.ispartofIEEE Transactions on Biomedical Engineering, 1989, vol. 36, núm. 11, p. 1137-1140.eng
dc.relation.urihttp://dx.doi.org/10.1109/10.40822-
dc.rights(c) IEEE, 1989cat
dc.sourceArticles publicats en revistes (Ciències Fisiològiques)-
dc.subject.classificationMesuramentcat
dc.subject.classificationErrors científicscat
dc.subject.otherBiomedical measurementeng
dc.subject.otherMeasurement errorseng
dc.subject.otherModelingeng
dc.subject.otherPneumodynamicseng
dc.titleA correction procedure for the asymmetry of differential pressure transducers in respiratory impedance measurementseng
dc.typeinfo:eu-repo/semantics/articlecat
dc.typeinfo:eu-repo/semantics/publishedVersion-
dc.identifier.idgrec35255cat
dc.rights.accessRightsinfo:eu-repo/semantics/openAccess-
Appears in Collections:Articles publicats en revistes (Ciències Fisiològiques)

Files in This Item:
File Description SizeFormat 
35255.pdf487.51 kBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.