Navajas Navarro, DanielFarré Ventura, RamonRotger i Estapé, Maria del MarPeslin, René2009-06-172009-06-1719880018-9294https://hdl.handle.net/2445/8686This paper presents a new respiratory impedance estimator to minimize the error due to breathing. Its practical reliability was evaluated in a simulation using realistic signals. These signals were generated by superposing pressure and flow records obtained in two conditions: 1) when applying forced oscillation to a resistance- inertance- elastance (RIE) mechanical model; 2) when healthy subjects breathed through the unexcited forced oscillation generator. Impedances computed (4-32 Hz) from the simulated signals with the new estimator resulted in a mean value which was scarcely biased by the added breathing (errors less than 1 percent in the mean R, I , and E ) and had a small variability (coefficients of variation of R, I, and E of 1.3, 3.5, and 9.6 percent, respectively). Our results suggest that the proposed estimator reduces the error in measurement of respiratory impedance without appreciable extracomputational cost.5 p.application/pdfeng(c) IEEE, 1988Errors científicsMesuramentBiomedical measurementMeasurement errorsPneumodynamicsA new estimator to minimize the error due to breathing in the measurement of respiratory impedanceinfo:eu-repo/semantics/article35251info:eu-repo/semantics/openAccess