Fernández Varea, José MaríaBoyer López, Helena2017-09-062017-09-062017-01https://hdl.handle.net/2445/115049Treballs Finals de Grau de Física, Facultat de Física, Universitat de Barcelona, Curs: 2017, Tutor: José M. Fernández-VareaSemiconductor detectors are routinely used in [gamma]-ray and x-ray spectrometry. Their characteriza- tion is crucial in order to perform quantitative analysis. Three steps are needed to this end: energy and FWHM calibration and determination of the full-energy peak effciency. In this TFG we have performed these steps for a HPGe detector. We confirmed the very high linearity, excellent energy resolution and high effciency of the studied spectrometer.5 p.application/pdfengcc-by-nc-nd (c) Boyer, 2017http://creativecommons.org/licenses/by-nc-nd/3.0/es/Detectors de radiacióSemiconductorsTreballs de fi de grauNuclear countersSemiconductorsBachelor's thesesCharacterization of semiconductor detectors for [gamma]-ray and x-ray spectrometryinfo:eu-repo/semantics/bachelorThesisinfo:eu-repo/semantics/openAccess