Title: Performance characteristics of the whole-body discovery IQ PET/CT system
Author: Reynés-Llompart, Gabriel
Gámez, Cristina
Romero-Zayas, Immaculada
Rodríguez-Bel, Laura
Vercher Conejero, José Luís
Martí-Climent, Josep M.
Keywords: Tumors
Diagnòstic per la imatge
Tomografia per emissió de positrons
Imatge corporal
Tumors
Diagnostic imaging
Positron emission tomography
Body image
Issue Date: Jul-2017
Publisher: The Society of Nuclear Medicine and Molecular Imaging
Abstract: The aim of this study was to assess the physical performance of a new PET/CT system, the Discovery IQ with 5-ring detector blocks. Methods: Performance was measured using the National Electrical Manufacturers Association NU2-2012 methodology. Image quality was extended by accounting for different acquisition parameters (lesion-to-background ratios [8:1, 4:1, and 2:1] and acquisition times) and reconstruction algorithms (VUE-point HD [VPHD], VPHD with point-spread-function modeling [VPHD-S], and Q.Clear). Tomographic reconstruction was also assessed using a Jaszczak phantom. Additionally, 30 patient lesions were analyzed to account for differences in lesion volume and SUV quantification between reconstruction algorithms. Results: Spatial resolution ranged from 4.2 mm at 1 cm to 8.5 mm at 20 cm. Sensitivity measured at the center and at 10 cm was 22.8 and 20.4 kps/kBq, respectively. The noise-equivalent counting rate peak was 124 kcps at 9.1 kBq/cm3 The scatter fraction was 36.2%. The accuracy of correction for count losses and randoms was 3.9%. In the image quality test, contrast recovery for VPHD, VPHD-S, and Q.Clear ranged from 18%, 18%, and 13%, respectively (hot contrast; 10-mm sphere diameter; ratio, 2:1), to 68%, 67%, and 81%, respectively (cold contrast; 37-mm sphere diameter; ratio, 8:1). Background variability ranged from 3.4%, 3.0%, and 2.1%, respectively (ratio, 2:1), to 5.5%, 4.8%, and 3.7%, respectively (ratio, 8:1). On Q.Clear reconstruction, the decrease in the penalty term (β) increased the contrast recovery coefficients and background variability. With the Jaszczak phantom, image quality increased overall when a reconstruction algorithm modeling the point-spread function was used, and use of Q.Clear increased the signal-to-noise ratio. Lesions analyzed using VPHD-S and Q.Clear had an SUVmean of 6.5 ± 3 and 7 ± 3, respectively (P < 0.01), and an SUVmax of 11 ± 4.8 and 12 ± 4, respectively (P < 0.01). No significant difference in mean lesion volume was found between algorithms. Conclusion: Among the various Discovery bismuth germanium oxide-based PET/CT scanners, the IQ with 5-ring detector blocks has the highest overall performance, with improved sensitivity and counting rate performance. Q.Clear reconstruction improves the PET image quality, with higher recovery coefficients and lower background variability.
Note: Reproducció del document publicat a: https://doi.org/10.2967/jnumed.116.185561
It is part of: Journal of Nuclear Medicine, 2017, vol. 58, num. 7, p. 1155-1161
URI: http://hdl.handle.net/2445/132887
Related resource: https://doi.org/10.2967/jnumed.116.185561
ISSN: 0161-5505
Appears in Collections:Articles publicats en revistes (Ciències Clíniques)
Articles publicats en revistes (Institut d'lnvestigació Biomèdica de Bellvitge (IDIBELL))

Files in This Item:
File Description SizeFormat 
687557.pdf940.51 kBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.