Please use this identifier to cite or link to this item: http://hdl.handle.net/2445/130019
Title: Evaluation of synthesis conditions for plastic scintillation foils used to measure alpha- and beta-emitting radionuclides
Author: Merín, R.
Tarancón Sanz, Alex
Mitev, K.
Georgiev, S.
Bagán Navarro, Héctor
García Martínez, José Francisco
Keywords: Anàlisi radioquímica
Radioactivitat
Radiochemical analysis
Radioactivity
Issue Date: 17-Nov-2018
Publisher: Springer Verlag
Abstract: Plastic scintillation foils of polystyrene and polycarbonate with a thickness between 45 and 200 μm, have been produced using the solvent evaporation method. PSfoils presented a reproducible thickness (10-20%). PSfoils were characterized by the measurement of 36Cl or 241Am. For 36Cl spectrum is located at medium energies since not all energy is deposited in the scintillator and not all betas interact with the foils. For 241Am the efficiency values are very high and spectrum is a sharp peak located at high energies. 222Rn absorption (LD and K) and desorption capacities of the PSfoils have been also evaluated.
Note: Versió postprint del document publicat a: https://doi.org/10.1007/s10967-018-6341-z
It is part of: Journal of Radioanalytical and Nuclear Chemistry, 2018, vol. 319, num. 1, p. 135-145
URI: http://hdl.handle.net/2445/130019
Related resource: https://doi.org/10.1007/s10967-018-6341-z
ISSN: 0236-5731
Appears in Collections:Articles publicats en revistes (Enginyeria Química i Química Analítica)
Articles publicats en revistes (Institut de Recerca de l'Aigua (IdRA))

Files in This Item:
File Description SizeFormat 
685124.pdf187.12 kBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.