A new estimator to minimize the error due to breathing in the measurement of respiratory impedance
| dc.contributor.author | Navajas Navarro, Daniel | cat |
| dc.contributor.author | Farré Ventura, Ramon | cat |
| dc.contributor.author | Rotger i Estapé, Maria del Mar | cat |
| dc.contributor.author | Peslin, René | cat |
| dc.date.accessioned | 2009-06-17T08:40:18Z | |
| dc.date.available | 2009-06-17T08:40:18Z | |
| dc.date.issued | 1988 | cat |
| dc.description.abstract | This paper presents a new respiratory impedance estimator to minimize the error due to breathing. Its practical reliability was evaluated in a simulation using realistic signals. These signals were generated by superposing pressure and flow records obtained in two conditions: 1) when applying forced oscillation to a resistance- inertance- elastance (RIE) mechanical model; 2) when healthy subjects breathed through the unexcited forced oscillation generator. Impedances computed (4-32 Hz) from the simulated signals with the new estimator resulted in a mean value which was scarcely biased by the added breathing (errors less than 1 percent in the mean R, I , and E ) and had a small variability (coefficients of variation of R, I, and E of 1.3, 3.5, and 9.6 percent, respectively). Our results suggest that the proposed estimator reduces the error in measurement of respiratory impedance without appreciable extracomputational cost. | eng |
| dc.format.extent | 5 p. | cat |
| dc.format.mimetype | application/pdf | eng |
| dc.identifier.idgrec | 35251 | cat |
| dc.identifier.issn | 0018-9294 | cat |
| dc.identifier.uri | https://hdl.handle.net/2445/8686 | |
| dc.language.iso | eng | eng |
| dc.publisher | IEEE | cat |
| dc.relation.isformatof | Reproducció del document publicat a http://dx.doi.org/10.1109/10.8684 | cat |
| dc.relation.ispartof | IEEE Transactions on Biomedical Engineering, 1988, vol. 35, núm. 12, p. 1001-1005. | eng |
| dc.relation.requires | http://dx.doi.org/10.1109/10.8684 | |
| dc.rights | (c) IEEE, 1988 | eng |
| dc.rights.accessRights | info:eu-repo/semantics/openAccess | |
| dc.source | Articles publicats en revistes (Ciències Fisiològiques) | |
| dc.subject.classification | Errors científics | cat |
| dc.subject.classification | Mesurament | cat |
| dc.subject.other | Biomedical measurement | eng |
| dc.subject.other | Measurement errors | eng |
| dc.subject.other | Pneumodynamics | eng |
| dc.title | A new estimator to minimize the error due to breathing in the measurement of respiratory impedance | eng |
| dc.type | info:eu-repo/semantics/article | cat |