A new estimator to minimize the error due to breathing in the measurement of respiratory impedance

dc.contributor.authorNavajas Navarro, Danielcat
dc.contributor.authorFarré Ventura, Ramoncat
dc.contributor.authorRotger i Estapé, Maria del Marcat
dc.contributor.authorPeslin, Renécat
dc.date.accessioned2009-06-17T08:40:18Z
dc.date.available2009-06-17T08:40:18Z
dc.date.issued1988cat
dc.description.abstractThis paper presents a new respiratory impedance estimator to minimize the error due to breathing. Its practical reliability was evaluated in a simulation using realistic signals. These signals were generated by superposing pressure and flow records obtained in two conditions: 1) when applying forced oscillation to a resistance- inertance- elastance (RIE) mechanical model; 2) when healthy subjects breathed through the unexcited forced oscillation generator. Impedances computed (4-32 Hz) from the simulated signals with the new estimator resulted in a mean value which was scarcely biased by the added breathing (errors less than 1 percent in the mean R, I , and E ) and had a small variability (coefficients of variation of R, I, and E of 1.3, 3.5, and 9.6 percent, respectively). Our results suggest that the proposed estimator reduces the error in measurement of respiratory impedance without appreciable extracomputational cost.eng
dc.format.extent5 p.cat
dc.format.mimetypeapplication/pdfeng
dc.identifier.idgrec35251cat
dc.identifier.issn0018-9294cat
dc.identifier.urihttps://hdl.handle.net/2445/8686
dc.language.isoengeng
dc.publisherIEEEcat
dc.relation.isformatofReproducció del document publicat a http://dx.doi.org/10.1109/10.8684cat
dc.relation.ispartofIEEE Transactions on Biomedical Engineering, 1988, vol. 35, núm. 12, p. 1001-1005.eng
dc.relation.requireshttp://dx.doi.org/10.1109/10.8684
dc.rights(c) IEEE, 1988eng
dc.rights.accessRightsinfo:eu-repo/semantics/openAccess
dc.sourceArticles publicats en revistes (Ciències Fisiològiques)
dc.subject.classificationErrors científicscat
dc.subject.classificationMesuramentcat
dc.subject.otherBiomedical measurementeng
dc.subject.otherMeasurement errorseng
dc.subject.otherPneumodynamicseng
dc.titleA new estimator to minimize the error due to breathing in the measurement of respiratory impedanceeng
dc.typeinfo:eu-repo/semantics/articlecat