Document type

Conference object

Version

Accepted version

Publication date

All rights reserved

Please use this identifier to cite or link to this item: https://hdl.handle.net/2445/219969

Study of quality controls for stability check of the ROIs of a ketones mixture in different GC-IMS measurement campaigns

Journal Title

Director/Tutor

Additional contribution

Journal ISSN

Volume Title

Abstract

GC-IMS is a very good complementary technique to traditional GC-MS, that presents some advantages, but also, some disadvantages such as misalignments produced by many parameters affecting the equipment stability. The reproducibility of the measures has been studied in two different measurement campaigns with a set of automatized quality control parameters. Figures of merit from one region of interest present in the samples show that the saturation and asymmetry do not change between measurement campaigns, but the volume and area of the total ion spectra change. A correction of these changes between batches should be developed.

Citation

Citation

MALLAFRÉ MURO, Celia, et al. Study of quality controls for stability check of the ROIs of a ketones mixture in different GC-IMS measurement campaigns. Comunicació a: 2022 IEEE International Symposium on Olfaction and Electronic Nose (ISOEN). International Society for Olfaction and Chemical Sensing (ISOCS). Vol.  Aveiro, num. Portugal, pags. 29 de maig - 1 de juny 2022. [consulted: 15 of October of 2026]. Available at: https://hdl.handle.net/2445/219969

Export metadata

JSON - METS

Share record