Carregant...
Tipus de document
Treball de fi de grauData de publicació
Llicència de publicació
Si us plau utilitzeu sempre aquest identificador per citar o enllaçar aquest document: https://hdl.handle.net/2445/180826
Design of Surface Plasmon Resonance Devices for Biosensing
Títol de la revista
Autors
Director/Tutor
ISSN de la revista
Títol del volum
Recurs relacionat
Resum
A study of the effects of geometry of a SPR device based on a gold nanohole array for its use as a biosensor has been done. For this, the OptiFDTD software has been used. It has allowed to obtain the dependence between the position of the SPR dip and the period, the diameter of the hole and the thickness of the gold film. The simulations have been done for mediums with different refractive index to observe the effect of using different substances. Small shifts in the SPR peak have been observed caused by little changes in the refractive index. The results obtained are the expected ones, as there are many biosensing devices based on this principle.
Descripció
Treballs Finals de Grau de Física, Facultat de Física, Universitat de Barcelona, Curs: 2021, Tutor: Mauricio Moreno Sereno
Matèries (anglès)
Citació
Col·leccions
Citació
LATIF AKHTAR, Kamran. Design of Surface Plasmon Resonance Devices for Biosensing. [consulta: 24 de gener de 2026]. [Disponible a: https://hdl.handle.net/2445/180826]