On the need of in vivo verifications as quality control for intraoperative electron radiotherapy in breast cancer

Títol de la revista

Director/Tutor

ISSN de la revista

Títol del volum

Resum

Introduction: Intraoperative electron radiotherapy (IOERT) is a technique aiming to deliver radiotherapy during oncological surgery. In breast IOERT, the applicator and shielding disc placement are correlated with organs at risk (OAR) irradiation, in vivo verification of these parameters is scarcely reported. The aim of our study is to report and analyze possible causes of the misalignment using radiochromic films and compare our results to others reported in the bibliography. Methods: From November 2019 to April 2023, in vivo verifications were performed for 33 patients. IOERT was performed using a LIAC 10 MeV (Sordina, Italy) electron accelerator. We attached a radiochromic film to the upper side of the polytetrafluoroethylene cover of the shielding disc. The percentage of the irradiation area outside the disc was recorded and various parameters (applicator angulations, prescription depth, tumor location and breast size) were analyzed to find possible correlations. Results: For 29 patients, 20 Gy were prescribed while 10 Gy were prescribed to 4 patients. The average irradiated area outside the disc was 19% (0-56%) corresponding to a surface of 4.5 cm2 (0-17.4 cm2). The applicator of 5 cm was used for most of the patients. The mean prescription depth was 1.4 cm (0.5-2.5 cm). We found no correlation between the analyzed parameters and misalignment. Conclusion: This study confirms the presence and magnitude of the misalignments. We strongly recommend in vivo verifications as a quality check during IOERT procedures. The misalignment has no correlation with tumor localization parameters, so the solution could be based on technical improvements of the applicator.

Citació

Citació

CASES, Carla, OSES, Gabriela, HERREROS, Antonio, TARRATS ROSELL, Jordi, MORENO, Sara, MOLLÀ, Meritxell. On the need of in vivo verifications as quality control for intraoperative electron radiotherapy in breast cancer. _Clinical & Translational Oncology_. 2024. Vol. 26, núm. 7, pàgs. 1623-1629. [consulta: 20 de gener de 2026]. ISSN: 1699-048X. [Disponible a: https://hdl.handle.net/2445/217312]

Exportar metadades

JSON - METS

Compartir registre