The implications of particle energy and acidic media on gross alpha and gross beta determination using liquid scintillation

dc.contributor.authorZapata García, Daniel
dc.contributor.authorLlauradó i Tarragó, Montserrat
dc.contributor.authorRauret i Dalmau, Gemma
dc.date.accessioned2020-05-05T08:58:42Z
dc.date.available2020-05-05T08:58:42Z
dc.date.issued2011-12-16
dc.date.updated2020-05-05T08:58:42Z
dc.description.abstractThe interaction of humans with radioactivity present in the environment from natural and artificial sources necessitates an evaluation of its risk on human health. Gross alpha and gross beta activities can provide a rapid evaluation of the radioactive content of a sample and can be simultaneously determined by using liquid scintillation counters. However, calibration of the liquid scintillation counter is required and is affected by many factors, such as particle energy and the acidity of the media. This study investigates what effect the particle energy used for calibration has on misclassification and how to account for this misclassification in routine measurements. The variability in measurement produced by the final pH, as well as any acids used in sample treatment, was also studied. These results showed that the most commonly used acid for these types of analyses, HNO3, produced a high amount of misclassifications at very low pH. The results improved when HCl was used to adjust the sample to low pH.
dc.format.extent7 p.
dc.format.mimetypeapplication/pdf
dc.identifier.idgrec606597
dc.identifier.issn0969-8043
dc.identifier.urihttps://hdl.handle.net/2445/158641
dc.language.isoeng
dc.publisherElsevier Ltd
dc.relation.isformatofVersió postprint del document publicat a: https://doi.org/10.1016/j.apradiso.2011.12.018
dc.relation.ispartofApplied Radiation and Isotopes, 2011, vol. 70, num. 4, p. 705-711
dc.relation.urihttps://doi.org/10.1016/j.apradiso.2011.12.018
dc.rights(c) Elsevier Ltd, 2011
dc.rights.accessRightsinfo:eu-repo/semantics/openAccess
dc.sourceArticles publicats en revistes (Enginyeria Química i Química Analítica)
dc.subject.classificationRadó
dc.subject.classificationAigua potable
dc.subject.classificationCalibratge
dc.subject.classificationEscintil·ladors líquids
dc.subject.otherRadon
dc.subject.otherDrinking water
dc.subject.otherCalibration
dc.subject.otherLiquid scintillators
dc.titleThe implications of particle energy and acidic media on gross alpha and gross beta determination using liquid scintillation