Development of an equipment for real-time continuous monitoring of alpha and beta radioactivity in river water

dc.contributor.authorTarancón Sanz, Alex
dc.contributor.authorNovella, O.
dc.contributor.authorPujadas, M.
dc.contributor.authorBatlle, Montserrat
dc.contributor.authorCros, J.
dc.contributor.authorGarcía Martínez, José Francisco
dc.date.accessioned2024-06-21T15:44:35Z
dc.date.available2024-06-21T15:44:35Z
dc.date.issued2022-06-10
dc.date.updated2024-06-21T15:44:40Z
dc.description.abstractDifferent regulations require the monitoring of radioactivity in the environment (e.g., 2013/51/Euratom, Real Decreto 314/2016) to protect the environment and the population from abnormal radioactivity presence caused by natural reasons or discharges or accidents in nuclear installations. Nowadays, the monitoring of α- and β-emitting radionuclides is performed discontinuously in laboratories due to the difficulties in applying classical techniques to continuous measurements. This limits the number of samples that can be measured per day, produces high costs per analysis, and introduces a significant delay between the moment of contamination and when it is detected. Plastic scintillation microspheres (PSm) represent a new possibility for continuous measurements because water samples can flow through a bed of PSm connected to a pair of photomultipliers (PMTs), allowing continuous monitoring of the activity. This idea is the basis of the Waterrad detector, which can monitor radioactivity at environmental levels in river water. This paper describes the optimization of a detection cell containing PSm, a detection chamber as well as active and passive shielding. In its final set-up, the Waterrad detector presents a background signal of 0.23 (1) cps and detection efficiencies of 1.86(7)⋅10 5 cps⋅L⋅Bq 1 for 3H, 7.4(8)⋅10 3 cps⋅L⋅Bq 1 for 90Sr/90Y and 5.5(5)⋅10 3 cps⋅L⋅Bq 1 for 241Am. The detection limits in the optimum window for a counting time of 5 h were 490 Bq/L for 3H, 2.3 Bq/L for 90Sr/90Y and 3.0 Bq/L for 241Am. These values indicate that Waterrad can be used as an alarm detector for monitoring radioactivity in water at activity levels similar to those of environmental samples, making it suitable for water or waste surveillance involving a high frequency of measurements.
dc.format.extent1 p.
dc.format.mimetypeapplication/pdf
dc.identifier.idgrec725509
dc.identifier.issn0969-8043
dc.identifier.urihttps://hdl.handle.net/2445/213512
dc.language.isoeng
dc.publisherElsevier Ltd
dc.relation.isformatofReproducció del document publicat a:
dc.relation.ispartofApplied Radiation and Isotopes, 2022, vol. 187, p. 110322
dc.rightscc-by-nc-nd (c) Tarancón Sanz, Alex, et al., 2022
dc.rights.accessRightsinfo:eu-repo/semantics/openAccess
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/
dc.sourceArticles publicats en revistes (Enginyeria Química i Química Analítica)
dc.subject.classificationRadioactivitat
dc.subject.classificationDetectors
dc.subject.classificationAnàlisi de l'aigua
dc.subject.otherRadioactivity
dc.subject.otherDetectors
dc.subject.otherWater analysis
dc.titleDevelopment of an equipment for real-time continuous monitoring of alpha and beta radioactivity in river water
dc.typeinfo:eu-repo/semantics/article