Validation of the Guide XT software for patients with Parkinson’s Disease undergoing STN-DBS in Hospital Clínic of Barcelona
Journal Title
Authors
Director/Tutor
Journal ISSN
Volume Title
Related resource
Abstract
Parkison’s Disease is the second most common neurodegenerative disorder worldwide and its
prevalence is expected to grow as population ages. Deep brain stimulation is a well-established
surgical treatment used as an alternative when medication is not effective. This neurosurgical
procedure consists of the implantation of electrodes into specific targets within certain areas of the
brain and on the electrical stimulation of these targets through the application of constant or
intermittent electricity from an implantable pulse generator (IPG) to treat the movement disorders
associated with Parkinson’s Disease. Stimulation parameters setting is highly important in this
surgical treatment. Programming deep brain stimulation is a challenging and time-consuming
process since there is a huge number of possible parameters combinations.
Guide XT (Boston Scientific) is a 3-dimensional neuroanatomic visual software that automatically
detects the position of the electrodes in the anatomy of the patient. It enables the visualization of
the volume of tissue activated by simulating the stimulation field output. This can help to efficiently
determine the most appropriate settings for each patient. The objective of this project is to compare
the stimulation settings of the traditional trial-and-error method versus the Guide XT-assisted
selection of parameters, to determine software reliability as a supporting tool in deep brain
stimulation programming.
For the study, data from 37 patients who underwent bilateral subthalamic nucleus deep brain
stimulation was collected. Stimulation settings from the first programming session and from the one
month after the surgery visit was obtained. All patients were implanted with the Vercise Cartesia
directional leads and the comparison and statistical analysis with respect to the stimulation
parameters selected with Guide XT was performed in a jupyter notebook.
The highest value of the Cohen’s kappa coefficient used to assess inter-rater agreement for the left
side contact selection for those patients who did not present side effects in the first programming
session was k=0.26. Nevertheless, the percentage of correspondence between contacts selection
for both electrodes in the two programming sessions and selection based on software simulations
was over 50% in all cases. Thus, it was concluded that the software is a reliable tool to be used in
the clinical practice, even though a further analysis testing results in patients would be of interest
Description
Treballs Finals de Grau d'Enginyeria Biomèdica. Facultat de Medicina i Ciències de la Salut. Universitat de Barcelona. Curs: 2021-2022. Tutor/Director: Pedro Roldán Ramos.
Subject (English)
Citation
Citation
PÉREZ MIRANDA, Cristina María. Validation of the Guide XT software for patients with Parkinson’s Disease undergoing STN-DBS in Hospital Clínic of Barcelona. [consulted: 18 of August of 2026]. Available at: https://hdl.handle.net/2445/186669