Fluorescence Multi-Detection Device Using a Lensless Matrix Addressable microLED Array
| dc.contributor.author | Moro Moreno, Víctor | |
| dc.contributor.author | Canals Gil, Joan | |
| dc.contributor.author | Moreno Martín, Sergio | |
| dc.contributor.author | Bornemann, Steffen | |
| dc.contributor.author | Alonso Casanovas, Oscar | |
| dc.contributor.author | Waag, Andreas | |
| dc.contributor.author | Prades García, Juan Daniel | |
| dc.contributor.author | Diéguez Barrientos, Àngel | |
| dc.date.accessioned | 2025-04-23T07:44:20Z | |
| dc.date.available | 2025-04-23T07:44:20Z | |
| dc.date.issued | 2024-05-21 | |
| dc.date.updated | 2025-04-23T07:44:20Z | |
| dc.description.abstract | A Point-of-Care system for molecular diagnosis (PoC-MD) is described, combining GaN and CMOS chips. The device is a micro-system for fluorescence measurements, capable of analyzing both intensity and lifetime. It consists of a hybrid micro-structure based on a 32 × 32 matrix addressable GaN microLED array, with square LEDs of 50 µm edge length and 100 µm pitch, with an underneath wire bonded custom chip integrating their drivers and placed face-to-face to an array of 16 × 16 single-photon avalanche diodes (SPADs) CMOS. This approach replaces instrumentation based on lasers, bulky optical components, and discrete electronics with a full hybrid micro-system, enabling measurements on 32 × 32 spots. The reported system is suitable for long lifetime (>10 ns) fluorophores with a limit of detection ~1/4 µM. Proof-of-concept measurements of streptavidin conjugate Qdot™ 605 and Amino PEG Qdot™ 705 are demonstrated, along with the device ability to detect both fluorophores in the same measurement | |
| dc.format.extent | 17 p. | |
| dc.format.mimetype | application/pdf | |
| dc.identifier.idgrec | 748673 | |
| dc.identifier.issn | 2079-6374 | |
| dc.identifier.uri | https://hdl.handle.net/2445/220537 | |
| dc.language.iso | eng | |
| dc.publisher | MDPI | |
| dc.relation.isformatof | Reproducció del document publicat a: https://doi.org/10.3390/bios14060264 | |
| dc.relation.ispartof | Biosensors-Basel, 2024 | |
| dc.relation.uri | https://doi.org/10.3390/bios14060264 | |
| dc.rights | cc-by (c) Moro, V. et al., 2024 | |
| dc.rights.accessRights | info:eu-repo/semantics/openAccess | |
| dc.rights.uri | http://creativecommons.org/licenses/by/4.0/ | |
| dc.source | Articles publicats en revistes (Enginyeria Electrònica i Biomèdica) | |
| dc.subject.classification | Fluorescència | |
| dc.subject.classification | Díodes electroluminescents | |
| dc.subject.other | Fluorescence | |
| dc.subject.other | Light emitting diodes | |
| dc.title | Fluorescence Multi-Detection Device Using a Lensless Matrix Addressable microLED Array | |
| dc.type | info:eu-repo/semantics/article | |
| dc.type | info:eu-repo/semantics/publishedVersion |