The objective of this work is to characterize the intrinsic α background for different 3He neutron detectors in order to be able to use them in low-rate neutron experiments. Isolated from atmospheric neutrons, two detectors gathered data in order to detect the background radiation without any neutron interference. Isolation from atmospheric neutrons was achieved using an experimental setup involving a polyethylene moderator and a cadmium neutron absorber. The whole experiment was performed in the Canfranc Underground Laboratory. By using Pulse Shape Discrimination methods, the intrinsic α spectrum for each detector was obtained and the α detection rates were found to be (7.9±0.3) · 10−5 s−1 and (7.7±0.3) · 10−5 s−1 respectively. Comparing these results with the ones from similar detectors portrays the need to perform this characterization for every 3He detector.

Descripció

Treballs Finals de Grau de Física, Facultat de Física, Universitat de Barcelona, Curs: 2026, Tutors: Guillem Cortés, Assumpta Parreño

Citació

Citació

FONT CASUGAS, Nicolás. Characterization of Intrinsic Neutron Background in 3He detectors. [consulted: 21 of May of 2026]. Available at: https://hdl.handle.net/2445/226808