Simultaneous determination of iron and copper using screen-printed carbon electrodes by adsorptive stripping voltammetry with o-phenanthroline.

Títol de la revista

Director/Tutor

ISSN de la revista

Títol del volum

Resum

The simultaneous voltammetric determination of iron and copper was accomplished by an unmodified screen-printed carbon electrode based on the complexation with o-phenanthroline assisted by the reducing agent ferrocyanide. The detection limits were 3.74 and 0.34 µg L−1 for iron and copper, respectively, and a linear response was observed from 12.5 to 400 µg L−1 for iron and from 1.14 to 400 µg L−1 for copper. This method successfully avoids interferences between iron and copper while allowing the simultaneous determination of both low and high metal concentrations usually found in the environment. Finally, a certified wastewater reference material was successfully analyzed, confirming the precision and accuracy of the proposed method.

Matèries (anglès)

Citació

Citació

USTABASI, Gul S., et al. Simultaneous determination of iron and copper using screen-printed carbon electrodes by adsorptive stripping voltammetry with o-phenanthroline. Microchemical Journal. 2022. Vol. 179, núm. 107597. ISSN 0026-265X. [consulta: 10 de maig de 2026]. Disponible a: https://hdl.handle.net/2445/189554