dc.description.abstractPatients with cardiac implantable electrical devices should take precautions when exposed to electromagnetic fields. Possible interference as a result of proximity to electromagnets or electricity flow from electronic tools employed in clinical odontology remains controversial. The objective of this study was to examine in vitro the capacity of dental equipment to provoke electromagnetic interference in pacemakers and implantable cardioverter defibrillators. Six electronic dental instruments were tested on three implantable cardioverter defibrillators and three pacemakers from different manufacturers. A simulator model, submerged in physiological saline, with elements that reproduced life-size anatomic structures was used. The instruments were analyzed at differing distances and for different time periods of application. The dental instruments studied displayed significant differences in their capacity to trigger electromagnetic interference. Significant differences in the quantity of registered interference were observed with respect to the variables manufacturer, type of cardiac implant, and application distance but not with the variable time of application. The electronic dental equipment tested at a clinical application distance (20 cm) provoked only slight interference in the pacemakers and implantable cardioverter defibrillators employed, irrespective of manufacturer.
dc.format.extent8 p.
dc.format.mimetypeapplication/pdf
dc.identifier.idgrec652820
dc.identifier.issn0909-8836
dc.identifier.pmid25913780
dc.identifier.urihttps://hdl.handle.net/2445/108885
dc.language.isoeng
dc.publisherWiley
dc.relation.isformatofVersió postprint del document publicat a: https://doi.org/10.1111/eos.12183
dc.relation.ispartofEuropean Journal of Oral Sciences, 2015, vol. 123, num. 3, p. 194-201
dc.relation.urihttps://doi.org/10.1111/eos.12183
dc.rights(c) European Journal of Oral Sciences, 2015
dc.rights.accessRightsinfo:eu-repo/semantics/openAccess
dc.sourceArticles publicats en revistes (Odontoestomatologia)
dc.subject.classificationAparells i instruments odontològics
dc.subject.classificationCamps electromagnètics
dc.subject.classificationInterferència electromagnètica
dc.subject.classificationPolpa dental
dc.subject.classificationAparells i instruments quirúrgics
dc.subject.classificationImplants dentals
dc.subject.classificationDesfibril·ladors cardioversors implantables
dc.subject.otherDental instruments and apparatus
dc.subject.otherElectromagnetic fields
dc.subject.otherElectromagnetic interference
dc.subject.otherDental pulp
dc.subject.otherSurgical instruments and apparatus
dc.subject.otherDental implants
dc.subject.otherImplantable cardioverter-defibrillators
dc.titleCapacity of dental equipment to interfere with cardiac implantable electrical devices