Document type

Bachelor thesis

Publication date

Publication license

cc-by-nc-nd (c) Comas, 2025
Please use this identifier to cite or link to this item: https://hdl.handle.net/2445/222490

Use of a high-resolution 2D diode array for electronic brachytherapy quality control

Journal Title

Journal ISSN

Volume Title

Related resource

Abstract

Quality control in radiotherapy is essential to ensure treatment safety and accuracy. Our goal is to evaluate the efficacy of quality control of an electronic brachytherapy system, performed with a 2D diode array, and to propose new methods involving this tool. We have performed a variety of tests regarding system’s reproducibility, sensitivity and stability, and computed the gamma index and gamma pass rate to compare dose distributions. We have been able to determine our array’s resolution both in position and dose (at least 1 mm and 1%, respectively), and presented strategies to perform an electronic brachytherapy system’s quality control implementing this device

Description

Treballs Finals de Grau de Física, Facultat de Física, Universitat de Barcelona, Curs: 2025, Tutors: Carles Muñoz, Roser Sala

Citation

Citation

COMAS RODRÍGUEZ, Maria. Use of a high-resolution 2D diode array for electronic brachytherapy quality control. [consulted: 17 of June of 2026]. Available at: https://hdl.handle.net/2445/222490

Export metadata

JSON - METS

Share record