Use of a high-resolution 2D diode array for electronic brachytherapy quality control

dc.contributor.advisorMuñoz, Carles
dc.contributor.advisorSala Llonch, Roser
dc.contributor.authorComas Rodríguez, Maria
dc.date.accessioned2025-07-23T06:58:20Z
dc.date.available2025-07-23T06:58:20Z
dc.date.issued2025-01
dc.descriptionTreballs Finals de Grau de Física, Facultat de Física, Universitat de Barcelona, Curs: 2025, Tutors: Carles Muñoz, Roser Salaca
dc.description.abstractQuality control in radiotherapy is essential to ensure treatment safety and accuracy. Our goal is to evaluate the efficacy of quality control of an electronic brachytherapy system, performed with a 2D diode array, and to propose new methods involving this tool. We have performed a variety of tests regarding system’s reproducibility, sensitivity and stability, and computed the gamma index and gamma pass rate to compare dose distributions. We have been able to determine our array’s resolution both in position and dose (at least 1 mm and 1%, respectively), and presented strategies to perform an electronic brachytherapy system’s quality control implementing this deviceca
dc.format.extent7 p.
dc.format.mimetypeapplication/pdf
dc.identifier.urihttps://hdl.handle.net/2445/222490
dc.language.isoengca
dc.rightscc-by-nc-nd (c) Comas, 2025
dc.rights.accessRightsinfo:eu-repo/semantics/openAccessca
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/3.0/es/*
dc.sourceTreballs Finals de Grau (TFG) - Física
dc.subject.classificationBraquiteràpiacat
dc.subject.classificationRaigs Xcat
dc.subject.classificationTreballs de fi de graucat
dc.subject.otherRadioisotope brachytherapyeng
dc.subject.otherX-rayseng
dc.subject.otherBachelor's theseseng
dc.titleUse of a high-resolution 2D diode array for electronic brachytherapy quality controleng