Zn-Mg and Zn-Cu alloys for stenting applications: From nanoscale mechanical characterization to in vitro degradation and biocompatibility

dc.contributor.authorGarcía Mintegui, Claudia
dc.contributor.authorCatalina Córdoba, Laura
dc.contributor.authorBuxadera-Palomero, Judit
dc.contributor.authorMarquina, Marquina
dc.contributor.authorJiménez-Piqué, Emilio
dc.contributor.authorGinebra, Maria-Pau
dc.contributor.authorCortina, José Luis
dc.contributor.authorPegueroles, Marta
dc.date.accessioned2022-01-14T11:12:49Z
dc.date.available2022-01-14T11:12:49Z
dc.date.issued2021-05-04
dc.descriptionReproducció del document publicat a: https://doi.org/10.1016/j.bioactmat.2021.04.015ca
dc.description.abstractIn the recent decades, zinc (Zn) and its alloys have been drawing attention as promising candidates for bioresorbable cardiovascular stents due to its degradation rate more suitable than magnesium (Mg) and iron (Fe) alloys. However, its mechanical properties need to be improved in order to meet the criteria for vascular stents. This work investigates the mechanical properties, biodegradability and biocompatibility of Zn-Mg and Zn-Cu alloys in order to determine a proper alloy composition for optimal stent performance. Nanoindentation measurements are performed to characterize the mechanical properties at the nanoscale as a function of the Zn microstructure variations induced by alloying. The biodegradation mechanisms are discussed and correlated to microstructure, mechanical performance and bacterial/cell response. Addition of Mg or Cu alloying elements refined the microstructure of Zn and enhanced yield strength (YS) and ultimate tensile strength (UTS) proportional to the volume fraction of secondary phases. Zn-1Mg showed the higher YS and UTS and better performance in terms of degradation stability in Hanks’ solution. Zn-Cu alloys presented an antibacterial effect for S. aureus controlled by diffusion mechanisms and by contact. Biocompatibility was dependent on the degradation rate and the nature of the corrosion products.ca
dc.description.sponsorshipFinancial support was received from Spanish Government, MINECO/FEDER, (RTI2018-098075-B-C21) and the Agency for Administration of University and Research Grants of the Government of Catalonia (2017SGR-1165). L.C.C. thanks COFUND scheme (GA 712754) and SEV-2014-0425 (2015–2019) for the financial support. Support for the research of M-P.G. was received through the prize “ICREA Academia” for excellence in research, funded by the Generalitat de Catalunya. Authors acknowledge Dr. Daniel Rodríguez-Rius for helping in the corrosion studies and measurement setup.ca
dc.format.extent17 p.
dc.format.mimetypeapplication/pdf
dc.identifier.citationBioactive Materials. 6 (12): 4430-4446ca
dc.identifier.issn2452-199X
dc.identifier.pmid34027233
dc.identifier.urihttps://hdl.handle.net/2445/182381
dc.language.isoengca
dc.publisherKeAi Communications Co. Ltd.ca
dc.relation.ispartofBioactive Materials, 2021, vol. 6, num. 12, p. 4430-4446
dc.relation.urihttps://doi.org/10.1016/j.bioactmat.2021.04.015
dc.rightscc by-nc-nd (c) García Mintegui, Claudia et al, 2021
dc.rights.accessRightsinfo:eu-repo/semantics/openAccessca
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/3.0/es/*
dc.sourceArticles publicats en revistes (Institut de Bioenginyeria de Catalunya (IBEC))
dc.subject.classificationZinc
dc.subject.classificationBiocompatibilitat
dc.subject.otherZinc
dc.subject.otherBiocompatibility
dc.titleZn-Mg and Zn-Cu alloys for stenting applications: From nanoscale mechanical characterization to in vitro degradation and biocompatibilityca
dc.typeinfo:eu-repo/semantics/articleca
dc.typeinfo:eu-repo/semantics/publishedVersion

Fitxers

Paquet original

Mostrant 1 - 1 de 1
Carregant...
Miniatura
Nom:
1-s2.0-S2452199X21001845-main.pdf
Mida:
25.23 MB
Format:
Adobe Portable Document Format
Descripció: