Analog circuit for real-time computation of respiratory mechanical impedance in sleep studies

Journal Title

Director/Tutor

Journal ISSN

Volume Title

Abstract

The aim of this work was to develop a low-cost circuit for real-time analog computation of the respiratory mechanical impedance in sleep studies. The practical performance of the circuit was tested in six patients with obstructive sleep apnea. The impedance signal provided by the analog circuit was compared with the impedance calculated simultaneously with a conventional computerized system. We concluded that the low-cost analog circuit developed could be a useful tool for facilitating the real-time assessment of airway obstruction in routine sleep studies.

Citation

Citation

FARRÉ VENTURA, Ramon, et al. Analog circuit for real-time computation of respiratory mechanical impedance in sleep studies. IEEE Transactions on Biomedical Engineering. 1997. Vol. 44, num. 11, pags. 1156-1159. ISSN 0018-9294. [consulted: 19 of August of 2026]. Available at: https://hdl.handle.net/2445/8685