Aortic perforation by active-fixation atrial pacing lead: an unusual but serious complication

dc.contributor.authorMarco, Andrea Di
dc.contributor.authorNuñez, Elaine
dc.contributor.authorOsorio, Karina
dc.contributor.authorDallaglio, Paolo D.
dc.contributor.authorAnguera Camós, Ignasi
dc.contributor.authorToscano, Jacobo
dc.contributor.authorSabaté, Xavier, 1950-
dc.contributor.authorCequier Fillat, Àngel R.
dc.date.accessioned2016-03-17T16:45:23Z
dc.date.available2016-03-17T16:45:23Z
dc.date.issued2014-06-01
dc.date.updated2016-03-17T16:45:28Z
dc.description.abstractPerforation of a cardiac chamber is an infrequent but serious sequela of pacemaker lead implantation. An even rarer event is the perforation of the aorta by a protruding right atrial wire. We present here the first case in the medical literature of aortic perforation as a sequela to the implantation of a cardiac resynchronization therapy defibrillator. The patient was a 54-year-old man with idiopathic dilated cardiomyopathy who underwent the implantation of a defibrillator, with no apparent sequelae. Six hours after the procedure, he experienced cardiac tamponade and required urgent open-chest surgery. The pericardial effusion was found to be caused by mechanical friction of a protruding right atrial wire on the aortic root. The aortic root and the atrial wall were both repaired with Prolene suture, which achieved complete control of the bleeding. There was no need to reposition the atrial wire. The patient had a good postoperative recovery.
dc.format.extent2 p.
dc.format.mimetypeapplication/pdf
dc.identifier.idgrec646763
dc.identifier.issn0730-2347
dc.identifier.pmid24955055
dc.identifier.urihttps://hdl.handle.net/2445/96595
dc.language.isoeng
dc.publisherTexas Heart Institute
dc.relation.isformatofReproducció del document publicat a: http://dx.doi.org/10.14503/THIJ-13-3219
dc.relation.ispartofTexas Heart Institute Journal, 2014, vol. 41, num. 3, p. 327-328
dc.relation.urihttp://dx.doi.org/10.14503/THIJ-13-3219
dc.rights(c) Texas Heart Institute, 2014
dc.rights.accessRightsinfo:eu-repo/semantics/openAccess
dc.sourceArticles publicats en revistes (Ciències Clíniques)
dc.subject.classificationMalalties de l'aorta
dc.subject.classificationDesfibril·ladors cardioversors implantables
dc.subject.classificationMarcapassos
dc.subject.otherAorta diseases
dc.subject.otherImplantable cardioverter-defibrillators
dc.subject.otherCardiac pacemakers
dc.titleAortic perforation by active-fixation atrial pacing lead: an unusual but serious complication
dc.typeinfo:eu-repo/semantics/article