Undetected displacement of a subcutaneous implantable cardioverter-defibrillator lead

dc.contributor.authorApolo, José
dc.contributor.authorSan Antonio, Rodolfo
dc.contributor.authorMont Girbau, Lluís
dc.contributor.authorTolosana, José M. (José María)
dc.date.accessioned2020-06-01T21:16:22Z
dc.date.available2020-06-01T21:16:22Z
dc.date.issued2019-10-22
dc.date.updated2020-06-01T21:16:23Z
dc.description.abstractBackground: In recent years, subcutaneous implantable cardioverter-defibrillator (S-ICD) implants have progressively increased and have been shown to be safe and highly successful, affording low reintervention rates regardless of the technique used. Case summary: We present a case of S-ICD implantation in a patient diagnosed with idiopathic ventricular fibrillation. In the first follow-up consultation the patient showed appropriate detection parameters in the three configurations. However, chest X-ray revealed lead displacement with a tip migration from the manubrium area of the sternum to the xiphoid process. Discussion: This case highlights the importance of performing at least one chest X-ray during the first weeks after S-ICD implantation, allowing the detection of a problem such as lead displacement, which can lead to undersensing of ventricular arrhythmias or S-ICD oversensing.
dc.format.extent5 p.
dc.format.mimetypeapplication/pdf
dc.identifier.idgrec698389
dc.identifier.issn2514-2119
dc.identifier.urihttps://hdl.handle.net/2445/163506
dc.language.isoeng
dc.publisherOxford University Press
dc.relation.isformatofReproducció del document publicat a: https://doi.org/10.1093/ehjcr/ytz189
dc.relation.ispartofEuropean Heart Journal - Case Reports, 2019, vol. 3, num. 4, p. 1-5
dc.relation.urihttps://doi.org/10.1093/ehjcr/ytz189
dc.rightscc-by-nc (c) Apolo, José et al., 2019
dc.rights.accessRightsinfo:eu-repo/semantics/openAccess
dc.rights.urihttp://creativecommons.org/licenses/by-nc/3.0/es
dc.sourceArticles publicats en revistes (Medicina)
dc.subject.classificationRaigs X
dc.subject.classificationDesfibril·ladors cardioversors implantables
dc.subject.classificationFibril·lació ventricular
dc.subject.otherX-rays
dc.subject.otherImplantable cardioverter-defibrillators
dc.subject.otherVentricular fibrillation
dc.titleUndetected displacement of a subcutaneous implantable cardioverter-defibrillator lead