Carregant...
Miniatura

Tipus de document

Article

Versió

Versió publicada

Data de publicació

Tots els drets reservats

Si us plau utilitzeu sempre aquest identificador per citar o enllaçar aquest document: https://hdl.handle.net/2445/119529

Failure of ProTaper rotary Ni-Ti instruments used by undergraduate students

Títol de la revista

Director/Tutor

ISSN de la revista

Títol del volum

Resum

Objective: To evaluate the effect of number of uses, angle and radius of curvature and type of instrument on the fracture of ProTaper rotary instruments when used by undergraduate students. Study design: Three hundred and seventy-six molars, with a total of 1114 root canals, extracted were instrumented by undergraduate students using ProTaper instruments according to the manufacturer´s recommendations. When fracture occurred, data were collected concerning the number of uses, type of instrument, level of fracture, angle and radius of curvature. ANOVA test were used to determine the influence of type of instrument in the incidence of instrument fracture. Logistic regression model was used to determine the influence of number of uses, angle and radius of curvature in the incidence of instrument fracture. Significance was set at p< 0.05. Results: A total of 37 Ni-Ti rotary instruments fractured during the treatment. Fracture occurred in 9.84% (37/376) of the teeth treated and 3.32% of the canals prepared with Ni-Ti rotary instruments (37/1114). A decrease in the radius of curvature of the canal significantly increased the likelihood of fracture (p=0.0001). Instrument fracture significantly increased as the number of uses increased (p=0.0037). No significant differences were found between the 6 types of ProTaper instruments (p=0.8). A reduction in the angle of curvature did not produce a significant decrease in the incidence of instrument separation (p=0.08). Conclusions: The results of this study imply that instrument fracture is linked to radius of curvature and number of uses

Citació

Citació

GARCÍA FONT, Marc, DURAN-SINDREU, Fernando, MORELLO CASTRO, Sergio, MERCADÉ BELLIDO, Montserrat, BUENO MARTÍNEZ, Rufino, ROIG CAYÓN, Miguel. Failure of ProTaper rotary Ni-Ti instruments used by undergraduate students. _Journal of Clinical and Experimental Dentistry_. 2012. Vol. 4, núm. 4, pàgs. e199-e203. [consulta: 21 de gener de 2026]. ISSN: 1989-5488. [Disponible a: https://hdl.handle.net/2445/119529]

Exportar metadades

JSON - METS

Compartir registre