Study of the accuracy of a radial arterial pressure waveform cardiac output measurement device after cardiac surgery

dc.contributor.authorOrdoñez Rufat, Pilar
dc.contributor.authorMancho Fora, Nuria
dc.contributor.authorTebé, Cristian
dc.contributor.authorPolit Martinez, Victoria
dc.contributor.authorAbellan Lencina, Ricardo
dc.contributor.authorFernandez Alvarez, Joaquin
dc.contributor.authorLópez Delgado, Juan Carlos
dc.date.accessioned2023-02-23T16:21:48Z
dc.date.available2023-02-23T16:21:48Z
dc.date.issued2023-01-17
dc.date.updated2023-02-23T14:15:21Z
dc.description.abstractBackgroundLess invasive monitoring, such as radial arterial pulse contour analysis (ProAQT (R) sensor), represents an alternative when hemodynamic monitoring is necessary to guide postoperative management and invasive monitoring is not technically feasible. The aim of the study is to evaluate the accuracy of the ProAQT (R) sensor cardiac output measurements in comparison with Pulmonary Artery Catheter (PAC) during the postoperative course of patients who underwent cardiac surgery with cardiopulmonary bypass.Case presentationProspective observational study in a Surgical Intensive Care Unit of a tertiary university hospital. Ten patients with a mean age of 73.5 years were included. The main comorbidities were hypertension, diabetes, dyslipidemia and the preoperative left ejection fraction was 43.8 +/- 14.5%. Regarding the type of surgery, six patients underwent valve surgery, two underwent coronary artery bypass grafting and two underwent aortic surgery. The cardiac index measured simultaneously by the ProAQT (R) sensor was compared with the PAC. The parameters were evaluated at predefined time points during the early postoperative courses (6 h, 12 h, 24 h, 48 h and 72 h). The degree of agreement with the cardiac index between the PAC and the ProAQT (R) sensor along the time points was measured using the concordance correlation coefficient, Bland-Altman analysis, and four-quadrant plot. Sixty-three pairs of measurements were analyzed. We showed that measurements of cardiac index were slightly higher with PAC (beta = - 0.146, p-value = 0.094). The concordance correlation coefficient for the additive model of cardiac index was 0.64 (95% Confidence Interval: 0.36, 0.82), indicating a high concordance between both sensors. Bland-Altmann analysis showed a mean bias of 0.45 L center dot min(-1)center dot m(-2), limits of agreement from - 1.65 to 2.3 L center dot min(-1)center dot m(-2), and percentage of error was 82.5%. Four-quadrant plot of changes in cardiac index showed a good concordance rate (75%), which increases after applying the exclusion zone (87%).ConclusionsIn patients undergoing cardiac surgery, the ProAQT (R) sensor may be useful to monitor cardiac index during the postoperative period, especially when more invasive monitoring is not possible.
dc.format.extent9 p.
dc.format.mimetypeapplication/pdf
dc.identifier.issn1749-8090
dc.identifier.pmid36650554
dc.identifier.urihttps://hdl.handle.net/2445/194031
dc.language.isoeng
dc.publisherSpringer Science and Business Media LLC
dc.relation.isformatofReproducció del document publicat a: https://doi.org/10.1186/s13019-023-02128-1
dc.relation.ispartofJournal of Cardiothoracic Surgery, 2023, vol. 18, num. 1, p. 32
dc.relation.urihttps://doi.org/10.1186/s13019-023-02128-1
dc.rightscc by (c) Ordoñez Rufat, Pilar et al., 2023
dc.rights.accessRightsinfo:eu-repo/semantics/openAccess
dc.rights.urihttp://creativecommons.org/licenses/by/3.0/es/*
dc.sourceArticles publicats en revistes (Institut d'lnvestigació Biomèdica de Bellvitge (IDIBELL))
dc.subject.classificationCirurgia cardiovascular
dc.subject.classificationHemodinàmica
dc.subject.classificationCura postoperatòria
dc.subject.classificationMonitoratge de pacients
dc.subject.otherCardiovascular surgery
dc.subject.otherHemodynamics
dc.subject.otherPostoperative care
dc.subject.otherPatient monitoring
dc.titleStudy of the accuracy of a radial arterial pressure waveform cardiac output measurement device after cardiac surgery
dc.typeinfo:eu-repo/semantics/article
dc.typeinfo:eu-repo/semantics/publishedVersion

Fitxers

Paquet original

Mostrant 1 - 1 de 1
Carregant...
Miniatura
Nom:
s13019-023-02128-1.pdf
Mida:
1.05 MB
Format:
Adobe Portable Document Format