Catheter tip distensibility substantially influences the aspiration force of thrombectomy devices

Journal Title

Director/Tutor

Journal ISSN

Volume Title

Abstract

Background A direct aspiration first pass thrombectomy (ADAPT) is a fast-growing technique for which a broad catalog of catheters that provide a wide range of aspiration forces can be used. We aimed to characterize different catheters' aspiration performance on stiff clots in an in vitro vascular model. We hypothesized that labeled catheter inner diameter (labeled-ID) is not the only parameter that affects the aspiration force (asp-F) and that thrombus–catheter tip interaction and distensibility also play a major role. Methods We designed an experimental setup consisting of a 3D-printed carotid artery immersed in a water deposit. We measured asp-F and distensibility of catheter tips when performing ADAPT on a stiff clot analog larger than catheter labeled-ID. Correlations between asp-F, catheter ID, and tip distensibility were statistically assessed. Results Experimental asp-F and catheter labeled-ID were correlated (r=0.9601; P<0.01). The relative difference between experimental and theoretical asp-F (obtained by the product of the tip’s section area by the vacuum pressure) correlated with tip’s distensibility (r=0.9050; P<0.01), evidencing that ADAPT performance is highly influenced by catheter tip shape-adaptability to the clot and that the effective ID (eff-ID) may differ from the labeled-ID specified by manufacturers. Eff-ID showed the highest correlation with experimental asp-F (r=0.9944; P<0.01), confirming that eff-ID rather than labeled-ID should be considered to better estimate the device efficiency. Conclusions Catheter tip distensibility can induce a significant impact on ADAPT performance when retrieving a stiff clot larger than the device ID. Our findings might contribute to optimizing thrombectomy strategies and the design of novel aspiration catheters.

Subject (English)

Citation

Citation

LI, Jun, et al. Catheter tip distensibility substantially influences the aspiration force of thrombectomy devices. Journal of Neurointerventional Surgery. 2021. Vol. 1-5. ISSN 1759-8478. [consulted: 17 of August of 2026]. Available at: https://hdl.handle.net/2445/219445

Export metadata

JSON - METS

Share record