Reproducibility of optical coherence tomography airway imaging
Títol de la revista
Director/Tutor
ISSN de la revista
Títol del volum
Recurs relacionat
Resum
Optical coherence tomography (OCT) is a promising imaging technique to evaluate small airway remodeling. However, the short-term insertion-reinsertion reproducibility of OCT for evaluating the same bronchial pathway has yet to be established. We evaluated 74 OCT data sets from 38 current or former smokers twice within a single imaging session. Although the overall insertion-reinsertion airway wall thickness (WT) measurement coefficient of variation (CV) was moderate at 12%, much of the variability between repeat imaging was attributed to the observer; CV for repeated measurements of the same airway (intra-observer CV) was 9%. Therefore, reproducibility may be improved by introduction of automated analysis approaches suggesting that OCT has potential to be an in-vivo method for evaluating airway remodeling in future longitudinal and intervention studies. (C) 2015 Optical Society of America
Matèries
Matèries (anglès)
Citació
Citació
KIRBY, Miranda, OHTANI, Keishi, NICKENS, Taylor, LOPEZ LISBONA, Rosa maria, LEE, Anthony m. d., SHAIPANICH, Tawimas, LANE, Pierre, MACAULAY, Calum, LAM, Stephen, COXSON, Harvey o.. Reproducibility of optical coherence tomography airway imaging. _Biomedical Optics Express_. 2015. Vol. 6, núm. 11, pàgs. 4365-4377. [consulta: 22 de febrer de 2026]. [Disponible a: https://hdl.handle.net/2445/124213]