Is the re-use of sterilized implant abutments safe enough? (Implant abutment safety)
Títol de la revista
Director/Tutor
ISSN de la revista
Títol del volum
Recurs relacionat
Resum
BACKGROUND: The reuse of implant healing abutments is common in dental practice. Effective elimination of bacteria and viruses is accomplished by conventional sterilization. The aim of this work was to explore the eventual survival of microorganisms on sterilized healing abutments and to rule out the presence of transmissible organic material after standard procedures. MATERIAL AND METHODS: A total of 55 healing abutments previously used in patients will be washed and sterilized in a steam autoclave at 121 C for 15 min. Each healing abutment will be cultured in Brain Heart Infusion broth (BHI) under strict aseptic conditions. Besides, two control groups will be included: one of 3 unused healing abutments, and the other of just medium. After 10 days at 37°C under a 5% CO2 100 µl of the broth will be plated on solid media (Brain Infusion Agar, BHIA) and Columbia Blood agar to test for sterility. The remaining volume will be centrifuged, the sediment fixed, and a Gram stain performed to discard the presence of non-cultivable microorganisms. Moreover, to determine the presence of remaining organic material after the cleaning and sterilizing treatments, the bioburden will be determined by measuring total organic carbon (TOC) in another 10 previously used healing abutments, cleaned and sterilized, that will be submerged in Milli-Q water and sonicated. RESULTS: No bacterial growth was detected on any of the 58 cultured abutments, indicating that the sterilization was completely satisfactory in terms of removal of live bacteria or spores. Nevertheless, significant amounts of organic carbon may still be recovered (up to 125,31 µg/abutment) after they have been sterilized. CONCLUSIONS: Significant amounts of the bioburden remained adhered to the surfaces in spite of the cleaning and sterilization procedures. Taking into account our results and data from other authors, the presence of infectious particles on the reused healing abutments such as prions cannot be ruled out.
Matèries
Matèries (anglès)
Citació
Citació
SÁNCHEZ GARCÉS, Ma. ángeles, VINUESA AUMEDES, Teresa, CIURANA, Joan, VIÑAS, Miquel. Is the re-use of sterilized implant abutments safe enough? (Implant abutment safety). _Medicina Oral_. Patología Oral y Cirugía Bucal. Vol. 2019, núm. 24, pàgs. 5. [consulta: 21 de gener de 2026]. ISSN: 1698-4447. [Disponible a: https://hdl.handle.net/2445/171335]