Development and Validation of a New High-Performance Liquid Chromatography Method for the Simultaneous Quantification of Coenzyme Q10, Phosphatidylserine, and Vitamin C from a Cutting-Edge Liposomal Vehiculization

dc.contributor.authorRuiz García, Marta
dc.contributor.authorPérez Lozano, Pilar
dc.contributor.authorMercadé Frutos, Débora
dc.contributor.authorNardi Ricart, Anna
dc.contributor.authorSuñé Pou, Marc
dc.contributor.authorCano Sarabia, Mary
dc.contributor.authorGarcía Jimeno, Sonia
dc.contributor.authorSuñé i Negre, Josep M. (Josep Maria)
dc.contributor.authorMaspoch, Daniel
dc.contributor.authorGarcía Montoya, Encarna
dc.date.accessioned2020-05-12T14:06:09Z
dc.date.available2020-05-12T14:06:09Z
dc.date.issued2019-11-26
dc.date.updated2020-05-12T14:06:09Z
dc.description.abstractA high-performance liquid chromatography (HPLC) method was developed to simultaneously quantify three widely used active substances such as coenzyme Q10, phosphatidylserine, and vitamin C. This new method optimizes current timing and costs in the analyses of these three active substances. Additionally, since the analyzed compounds were encapsulated on a cutting-edge liposomal formulation, further processing was necessary to be developed prior to HPLC analyses. The technique was studied and adequately validated in accordance with the guidelines of the International Council for Harmonisation of Technical Requirements for Pharmaceuticals for Human Use (ICH) regarding selectivity, linearity, accuracy, precision, and robustness. After data treatment of results, linear regressions for all active substances showed an optimal linearity with a correlation coefficient of >0.999 in the concentration range between 70 to 130% of the liposomal formulation and less than a 3% relative standard deviation (RSD) in accuracy and precision.
dc.format.extent6 p.
dc.format.mimetypeapplication/pdf
dc.identifier.idgrec694683
dc.identifier.issn2470-1343
dc.identifier.pmid31788602
dc.identifier.urihttps://hdl.handle.net/2445/159770
dc.language.isoeng
dc.publisherAmerican Chemical Society
dc.relation.isformatofReproducció del document publicat a: https://doi.org/10.1021/acsomega.9b02456
dc.relation.ispartofACS Omega , 2019, vol. 4, num. 22, p. 19710-19715
dc.relation.urihttps://doi.org/10.1021/acsomega.9b02456
dc.rights(c) American Chemical Society, 2019
dc.rights.accessRightsinfo:eu-repo/semantics/openAccess
dc.sourceArticles publicats en revistes (Farmàcia, Tecnologia Farmacèutica i Fisicoquímica)
dc.subject.classificationCromatografia de líquids
dc.subject.classificationCoenzims
dc.subject.otherLiquid chromatography
dc.subject.otherCoenzymes
dc.titleDevelopment and Validation of a New High-Performance Liquid Chromatography Method for the Simultaneous Quantification of Coenzyme Q10, Phosphatidylserine, and Vitamin C from a Cutting-Edge Liposomal Vehiculization
dc.typeinfo:eu-repo/semantics/article
dc.typeinfo:eu-repo/semantics/publishedVersion

Fitxers

Paquet original

Mostrant 1 - 1 de 1
Carregant...
Miniatura
Nom:
694683.pdf
Mida:
920.62 KB
Format:
Adobe Portable Document Format