Effect of implantoplasty on fracture resistance and surface roughness of standard diameter dental implants
Títol de la revista
Director/Tutor
ISSN de la revista
Títol del volum
Recurs relacionat
Resum
Objective: To assess the effect of implantoplasty on the fracture resistance, surface roughness, and macroscopic morphology of standard diameter (4.1 mm) external connection dental implants.Materials and methods: An in vitro study was conducted in 20 screw-shaped titanium dental implants with an external connection. In 10 implants, the threads and surface were removed and polished with high-speed burs (implantoplasty), while the remaining 10 implants were used as controls. The final implant dimensions were recorded. The newly polished surface quality was assessed by scanning electron microscopy (SEM) and by 3D surface roughness analysis using a confocal laser microscope. Finally, all the implants were subjected to a mechanical pressure resistance test. A descriptive analysis of the data was made. Also, Student's t tests were employed to detect differences regarding the compression tests.Results: Implantoplasty was carried out for a mean time of 10 min and 48 s (standard deviation (SD) of 1 min 22 s). Macroscopically, the resulting surface had a smooth appearance, although small titanium shavings and silicon debris were present. The final surface roughness (Sa values 0.1 +/- 0.02 mu m) was significantly lower than that of the original (0.75 +/- 0.08 mu m Sa) (p = .005). There was minimal reduction in the implant's inner body diameter (0.19 +/- 0.03 mm), and no statistically significant differences were found between the test and control implants regarding the maximum resistance force (896 vs 880 N, respectively).Conclusions: Implantoplasty, although technically demanding and time-consuming, does not seem to significantly alter fracture resistance of standard diameter external connection implants. A smooth surface with Sa values below 0.1 mu m can be obtained through the use of silicon polishers. A larger sample is required to confirm that implantoplasty does not significantly affect the maximum resistance force of standard diameter external -connection implants.
Matèries
Matèries (anglès)
Citació
Citació
COSTA-BERENGUER, Xavier, SÁNCHEZ TORRES, Alba, SANZ ALONSO, Mariano, BARBOSA DE FIGUEIREDO, Rui pedro, VALMASEDA CASTELLÓN, Eduardo, GARCÍA-GARCÍA, Marta. Effect of implantoplasty on fracture resistance and surface roughness of standard diameter dental implants. _Clinical Oral Implants Research_. 2017. Vol. 29, núm. 1, pàgs. 46-54. [consulta: 13 de gener de 2026]. ISSN: 0905-7161. [Disponible a: https://hdl.handle.net/2445/218292]